New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices (Q2262652)
From MaRDI portal
scientific article
| Language | Label | Description | Also known as |
|---|---|---|---|
| English | New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices |
scientific article |
Statements
New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices (English)
0 references
16 March 2015
0 references
Summary: Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5\%, corresponding to an improvement of 14\% compared to the standard estimator.