Experimental data for validation of new design of wave membrane blood pump (Q6716264)
From MaRDI portal
| This is the item page for this Wikibase entity, intended for internal use and editing purposes. Please use this page instead for the normal view: Experimental data for validation of new design of wave membrane blood pump |
Dataset published at Zenodo repository.
| Language | Label | Description | Also known as |
|---|---|---|---|
| English | Experimental data for validation of new design of wave membrane blood pump |
Dataset published at Zenodo repository. |
Statements
Matlab dataset that collects HQ in-vitro data (H: head pressure, Q: flow rate) storedin two separate columns. The operating point of the wave membrane is the following: frequency of oscillation f=44Hz, oscillation amplitude A=0.75 mm. These data have been produced in pump characterization bench at the developing company CorWave and they have been used for the validation of the numerical simulations in the J-shape design of the wave membrane blood pump.
0 references
16 June 2021
0 references